NEW MV 240 digital - The future of measurement microphones
1/2“ Microphone Preamplifier with integrated ADC
The innovative MV 240 digital is an all-in-one new development, which differs in every respect from all commercially available analog-digital solutions in terms of design, functionality and quality.
Outstanding features are, e.g:
- 32 bit data transmission
- Digital transmission of fixed calibrated sound pressure levels
- Full dynamics of the newly developed low noise ½ " Capsule
- With ¼"capsule -max.160dB sound pressure level
- dynamic range - 32 bit range - 186dB
- no separate power supply necessary
Exactness and simplified operation make it possible to optimize existing and find new, cost-saving applications (e.g. industry, environment, R&D, occupational health and safety, Study and training sector).
Existing disadvantages of previous AD solutions could be overcome, e.g:
- time-consuming and personnel-intensive installation and operation,
- Errors due to complicated coordination of several electronic and software components,
- handling the electrical voltage as a coupler by the user,
- Measurable / audible USB artifacts,
- Insufficient sensitivity or limited frequency range, etc.
Customer benefits of the new MV 240:
Only one measuring range for the entire dynamic range, i.e. no range switching necessary,
- direct digital sound pressure conversion (with exact sensor-ADU adjustment to exclude common sources of error),
- Sampling frequency up to 192kHz is supported,
- no measurable/audible USB artifacts, USB class compatible (USB 2/3), no additional driver necessary,
- User independence from special manufacturer software, possibility of using commercially free available measurement software,
- internally generated capsule polarization voltage 200V,
- Minimal inherent noise due to optimum adaptation of the measuring capsule to the AD converter,
- Compatible with all common OS, such as Windows® (XP,Vista, 7, 8, 8.1, 10),Linux, Mac OS®, ioS®),
- Electrical voltage in the system is irrelevant to the user due to direct digital sound pressure conversion.
Another major advantage is a test generator integrated in the microphone, which checks both the complete transmission and evaluation of a 32-bit data stream from the microphone via the interface to the measurement software on the PC, as well as a further integrated test generator, which generates a signal corresponding to the sound pressure of 1 Pascal to show the reference point in the respective measurement software (internal electrical voltage in the system is irrelevant, which increases measurement reliability).
Specification
- USB interface (4-pin LEMO®)
- power supply via USB
- capsule polarisation voltage:internally generated (200 V)
- 1 Hz to 80 kHz / 7 dBA to 160 dB
The AD conversion takes place in 2 level ranges (24bit each). An integrated processor combines both level ranges to a single-channel USB output signal/32bit (corresponds to already calibrated sound pressure values).
One channel of a 24 bit analog-to-digital converter converts the lower level range of the capsule signal and the other channel the upper level range. A following processor combines these two level ranges to the one-channel 32 bit output signal.
The output signal of the capsule is optimally adjusted to the ADC`s dynamic range by an impedance converter and a preamplifier with a very low noise level and THD.
The pre-amplifier converts sound pressure levels from 7 dB to 160 dB, a frequency range up to 80 kHz and generates internally a switchable 200 V polarisation voltage for the microphone capsule. For the use of the MV 240 no special hardware driver is necessary because it is detected automatically by different operating systems like Windows®, MacOS® or Linux.
The MV 240 is powered by the USB interface and generates internally
200 V polarisation voltage for the measurement microphone capsule that can be switched off when using electret measurement microphone cap.
www.microtechgefell.de